Maintenance of automatic external defibrillators (AED)
During a recent audit, a member discovered significant non-conformances pertaining to the storage and maintenance of AED. The consequences were potentially significant, and immediate re-assessment was undertaken of all such equipment at that member’s dive sites. AEDs should be available for immediate usage.
Our member’s audit process noted the following non-conformances:
![The battery had been removed or had been fitted incorrectly to the device. This was contrary to advice in the user manual for the item](https://imcaweb.blob.core.windows.net/wp-uploads/2015/11/IMCASF18-15-image1.png)
![The pads on the AED were out of date](https://imcaweb.blob.core.windows.net/wp-uploads/2015/11/IMCASF18-15-image2.png)
Our member considers this to be a significant issue. Inspection of other sites revealed that in some cases batteries were removed or detached from the units. Removing or detaching the batteries from these units will prevent the unit being used for an additional sixty seconds, and lessens the chances of survival for the patient.
Our member took the following actions:
- Confirmed that the battery was fitted in accordance with the manufacturer’s instructions;
- Confirmed that all equipment associated with the AED was ‘in-date’;
- Tested the readiness of AED in accordance with the manufacturer’s instructions;
- Appropriately trained personnel conducted a familiarization session on use of the AED with all of the crew;
- Ensured that the maintenance of the AED was added to the appropriate planned maintenance system.
Safety Event
Published: 19 November 2015
Download: IMCA SF 18/15
IMCA Safety Flashes
Submit a Report
IMCA Safety Flashes summarise key safety matters and incidents, allowing lessons to be more easily learnt for the benefit of all. The effectiveness of the IMCA Safety Flash system depends on Members sharing information and so avoiding repeat incidents. Please consider adding [email protected] to your internal distribution list for safety alerts or manually submitting information on incidents you consider may be relevant. All information is anonymised or sanitised, as appropriate.
IMCA’s store terms and conditions (https://www.imca-int.com/legal-notices/terms/) apply to all downloads from IMCA’s website, including this document.
IMCA makes every effort to ensure the accuracy and reliability of the data contained in the documents it publishes, but IMCA shall not be liable for any guidance and/or recommendation and/or statement herein contained. The information contained in this document does not fulfil or replace any individual’s or Member's legal, regulatory or other duties or obligations in respect of their operations. Individuals and Members remain solely responsible for the safe, lawful and proper conduct of their operations.